871 Laser Wavelength Meter

Optical Wavelength Meter by Bristol Instruments

Note: Your Quotation Request will be directed to Bristol Instruments.

The 871 Series Laser Wavelength Meter from Bristol Instruments operates from 375 to 1700 nm measuring absolute wavelength of both pulsed and CW lasers. The device can measure wavelength to an accuracy as high as ± 0.75 parts per million. The system calibrates automatically with a built-in wavelength standard. The calibration process takes less than two seconds, and can be set to activate at a chosen time interval, or based on the system’s internal thermal conditions. It has fastest sustained measurement rate of 1 kHz resulting in time resolution of 1 ms. The meter requires an input as low as 20 nJ (20 µW at 1 kHz). The wavelength meter is available in a system that measures 89 mm x 432 mm x 381 mm and weighs 7.65 kg.

The 871 Laser Wavelength Meter is available in two broad wavelength configurations, VIS (375 - 1100 nm) and NIR (650 - 1700 nm). A pre-aligned fiber-optic input ensures optimum alignment resulting in uncompromised accuracy. The system operates with a PC running under Windows, via a USB or Ethernet interface. Software is provided to control measurement parameters and to report data, or the system can become part of an experiment using a library of commands for custom or LabVIEW programming.

 Product Specifications

    Product Details

    • Part Number :
      871 Laser Wavelength Meter
    • Manufacturer :
      Bristol Instruments
    • Description :
      Fast & Reliable Laser Wavelength Meter to measure the wavelength of pulsed/CW lasers

    Fiber Parameters

    • Fiber Core Diameter :
      62.5 µm
    • Fiber Mode :
      Multi-Mode, Single Mode

    Applications

    • Application :
      Browser-based display application

    Connections & Interface

    • Connector :
      FC/UPC
    • Interface :
      Ethernet, RS-422, USB, USB 2.0

    Temperature

    • Operating Temperature display :
      15 to 30 Degree C
    • Storage Temperature :
      -10 to 70 Degree C

    General Parameters

    • Operation Mode :
      CW Lasers, Pulsed Lasers
    • Wavelength Range :
      375 to 1700 nm(NIR, VIS)
    • Wavelength Display Unit :
      THz, GHz, µm, cm-1, nm
    • Wavelength accuracy :
      ±1 ppb(± 300 MHz at 1000 nm)
    • Measurement Range :
      500 Hz to 1 kHz
    • Bandwidth :
      10 GHz
    • Input Power Range :
      0.2 µJ (375 nm, 700 nm, 650 nm, 1100 nm, 1600 nm), 0.5 µJ (1100 nm)
    • Power Requirements :
      90 - 264 VAC, 47 - 63 Hz, 50 VA max
    • Pressure :
      500 - 900 mm Hg
    • Relative Humidity :
      90% R.H. at + 40°C (no condensation)
    • Repetition Rate :
      500 Hz( VIS version) to 1 kHz(NIR version)
    • Warm up time :
      15 Minutes
    • Wavelength Calibration :
      Automatic with built-in wavelength standard
    • Wavelength Display Resolution :
      8 digits Display Resolution
    • Wavelength Repeatability :
      ± 0.1 ppm
    • Note :
      NIR & VIS Bands, Minimum Optical Input Signal: 0.2 µJ

    Physical Properties

    • Dimensions :
      3.5" x 17"” x 15" (89 mm x 432 mm x 381 mm) (H x W x L)
    • Weight :
      17 lbs (7.65 kg)